Automatic Cleveland Open Cup Flash Point Analyzer

Model: KR-K800

Cleveland open cup flash point tester

This automatic open cup flash ignition point tester is manufactured according to GB267 “Petroleum product flash point and ignition point measurement method (open cup method)”, GB3536 “Petroleum product flash point and ignition point measurement method (Cleveland open cup method)” standard method, It is an ideal instrument for performing flash point testing in petroleum, chemical, electric power, environmental protection, railway, scientific research and other fields.

 

Main Features:

1. Adopt high-speed signal processor control, adaptive PID control algorithm, and automatically adjust the heating curve according to standard requirements.

2. Built-in atmospheric pressure automatic detection chip, automatic measurement of atmospheric pressure, automatic correction of flash point value.

3. Data storage function, can store 300 results, and check the history at any time.

4. Differential detection, automatically correct system deviation and stop detecting and alarming when the temperature exceeds the value.

5. The instrument automatically completes lifting, stirring, ignition, detection, alarm, cooling, and printing.

6. Safe airless source, platinum alloy electric heating wire ignition mode, safe, stable, reliable, high test accuracy.

 

Technical parameters of flash point analyzer:

Display 10 inch 65K(resolution 800*600) Chinese and English touch screen
Measurement method Thermocouple differential detection
Measuring range room temperature to 400 ° C
Repeatability ±4℃
Printer High speed thermal printer, 56 mm paper width
Ignition method automatic electronic ignition
Temperature detection imported precision platinum resistance
Self-test function printer, igniter, solenoid valve, detection arm lift, ignition rod sweep, etc.
Dimensions 450*280*290mm
Power source AC220V±10%
Weight about 15 kg

 