HFRR High Frequency Reciprocating Rig Equipment ASTM D6079

Model: KR-HFRR80     

Standard: ASTM D6079

HFRR Test Machine

High Frequency Reciprocating Rig Equipment Features and advantages:

1. Stroke of the testing machine: the exciter stroke/mm: 1±0.02, and can be set between 20μm-2mm; the reflective linear photoelectric encoder (grating ruler) and the reading grating sensor are used to control the stroke. During the test, the real-time stroke actual value can be observed in the software operation platform; it has a stroke self-checking system; it has a high-frequency reciprocating limit protection device.

2. The test machine can realize the lubricity evaluation of vehicle diesel, vehicle gasoline, ethanol gasoline and methanol gasoline without replacing any parts. It only needs to select the corresponding test method in the control software to carry out and test Complete the lubricity evaluation of diesel and gasoline. When evaluating the lubricity of gasoline, the operation steps for evaluating the lubricity of diesel oil are the same.

3. After the test, the freckle pictures can be directly imported into the software to form a complete test report with the freckle pictures and various parameters.
4. The microscope is equipped with a separate imaging screen and is connected with the camera of the electronic imaging system; the pixel of the microscope camera photo is 5 million. In addition, it has three sets of observation systems, such as visual microscope and electronic imaging.

5. The main box of the system testing machine and the constant temperature and humidity controller are connected in a split type to avoid resonance; the constant temperature and humidity controller is a separate instrument unit, and its characteristic is that after setting the ambient temperature and humidity, the set value can be Carry out automatic adjustment, and ensure that the temperature and humidity values in the whole test process conform to the steady-state error value of its set value.

The main technical parameters
1. High frequency oscillation frequency: 50+ 1Hz, and can be set from 10 to 200Hz
2. Oscillation stroke: 1+0.02mm, and it can be set between 20 μm and 2.0mm
3. Test oil temperature:
3.1 Diesel lubricity evaluation 60+2℃, and can be set from room temperature to 150℃
3.2 The evaluation of gasoline lubricity can be adjusted from room temperature to 150 ℃ according to the test requirements
4. Test load: 200 ± 1g, and can be set between 0.1 and 1Kg
5. Oil sample volume:
5.1 The lubricity of diesel oil is rated as 2.0+ 0.2mL;
5.2 The lubricity of gasoline is evaluated as 6.0 ± 0.2 mL
6. Oil box surface area: 600 soil 100mm2 special oil box fixture
7. Test time: 75+0.1min, the test will automatically stop when the time is up
8. Wear scar measurement system: metallographic microscope, with electronic imaging system, can use operating software to measure wear scar and add it to the test report; equipped with imported original X\Y axis reader; 5 million pixel camera.
9. The temperature of the test chamber: the temperature control range is 0-50 °C, and the accuracy is 0.1 °C
10. The humidity of the test chamber: can be set within the range of 5 to 95% RH
11. The ambient temperature of the instrument: 0 to 50 °C
12. The ambient humidity of the instrument: 5 to 95%RH
13. Test ball: diameter is 6mm, material is ISO 683-17-100Gr6 steel, ISO 3290-1 (metal ball) grade 28, should reach Rockwell hardness
Degree HRC is 58~66, surface finish Ra<0.05μm
14. Test piece: Test piece: processed from annealed ISO 683-17-100Gr6 steel bar to have Vickers hardness “HV30” of 190-210, and
Grinded and polished to surface finish Ra<0.02μm
15. Control capacity of compressor refrigeration constant temperature and humidity system:
15.1 Full braking to keep the temperature and humidity inside the main box constant, which has the advantages of high efficiency, strong reliability, and good environmental tolerance.
15.2 The temperature inside the box is 23±1℃, and the relative humidity is 53%±2%RH. The user can set it according to the standard requirements of the test method;
15.3 The device is connected with the main box to avoid resonance, and the temperature and humidity in the box can be adjusted to the set range in a short time;
15.4 This equipment meets the control requirements for ambient temperature and humidity in the ISO 12156.1-2018 test method standard.

Online Inquire