KR-CPF80 Fully Automatic Closed Cup Flash Point Analyzer

Model No. KR-CPF80

Flash point tester ASTM D93

Under specified conditions, when the oil is heated to a certain temperature, the mixture of oil vapor and surrounding air will flash upon contact with the flame. The lowest flash temperature is called the flash point value of the oil.

The KR-CPF80 fully automatic closed flash point tester is manufactured in accordance with the new standard method 1 of the national standard GB/T261-2021 “Determination of Flash Point by Binsky Martin Closed Cup Method” and the requirements of ASTMD93. It is applied in departments such as petroleum, chemical, power, environmental protection, railway, and scientific research, and is an ideal instrument for conducting closed flash point testing

 

Main features:

 

  1. Adopting a new high-speed signal processor control, adaptive PID control algorithm, automatically adjusting the heating curve according to standard requirements.
  2. Equipped with an automatic atmospheric pressure detection chip, atmospheric pressure is automatically measured, calibrated, and the flash point value is automatically corrected.
  3. The automatic measurement of lifting, stirring, ignition, detection, alarm, cooling, and printing instruments has been completed.
  4. Safe and gas free, platinum alloy electric heating wire ignition method, safe, stable, reliable, with high testing accuracy.
  5. differential detection, automatic correction of system deviation, automatic stop detection and alarm for temperature exceeding the limit.
  6. optional operations in both Chinese and English interfaces, instrument programs can be upgraded during national standard revisions.
  7. optional USB ports can export data to a computer (optional).
  8. comes with 12 sets of data for temperature point correction.
  9. Meet the new standard 261-A of GB/T261-2021 261-B 261-C and each degree test are optional.

 

Main Technical Parameters:

 

Display mode: 10.1-inch touchscreen display
Measurement method: Thermocouple differential detection formula
Display range: From minus 59.9 degrees to -400 degrees
Measurement range: Room temperature -400 ℃ (to determine the flash point at minus zero temperature, the sample needs to be frozen first)
Resolution 0.1℃
Repeatability: National standard: When the flash point value is<104 ℃, the allowable deviation is ± 1 ℃, and when the flash point value is ≥ 104 ℃, the allowable deviation is ± 2 ℃

 

International standard: When the flash point value is ≤ 110 ℃, the allowable deviation is ± 1 ℃, and when the flash point value is>110 ℃, the allowable deviation is ± 2 ℃

 

Heating rate: Complies with GB/T261-2021, step A, step B, step C, testing at each degree
printer: Low energy high-speed thermal printer, 56mm paper width
Temperature detection: Imported platinum resistor (Pt100 Ω)
Ignition method: No gas source electronic ignition
Self-checking function: Automatic diagnosis of instrument faults
Cooling method: Forced air cooling

 