描述
Description
The KR-5453C automatic ultraviolet fluorescence sulfur analyzer is a fully automatic total sulfur testing instrument. It integrates advantages of elemental analyzers from Thermo Fisher (USA), Mitsubishi (Japan), Analytik Jena (Germany), and achieves re-innovation after digestion and absorption of related technologies. The instrument features high efficiency, high precision, simple operation and reliable stability, reaching internationally advanced level.
Application Scope
It is used for the determination of total sulfur content in liquid products in petrochemical production, scientific research institutions, trading enterprises, customs, commodity inspection, quality supervision systems and other fields.
Compliance Standards
ASTM D5453, D6667, D7183, ISO 20846, GB/T 17930, GB/T 34100, GB/T 6324.10, GB/T 11060.8, GB/T 11141, SH/T 0689, SH/T 1820, etc.
Technical Specifications
Measuring range: 0.1 mg/kg~10000 mg/kg~%
Detection limit: 0.05 mg/kg
Applicable samples: solid, liquid and gas (autosampler optional)
PMT high voltage range: DC 400V~900V
Correlation coefficient of calibration curve: ≥0.999
Temperature control range & accuracy: ambient temperature~1300℃, ±1℃
Repeatability:
0.1<x≤1.0 mg/kg,±0.1 mg/kg
1<x≤10 mg/kg, r≤6%
10<x≤100 mg/kg,r≤2%
Above 100 mg/kg,r≤1%
Features
Operated on Windows platform. The whole process including analysis and data processing is controlled by computer, with real-time working status displayed. Test parameters and test results can be saved or printed. Historical data query and graph replay are supported for comparative analysis.
Adopts popular pulsed ultraviolet fluorescence analysis method with light source signal modulation and demodulation. It effectively avoids unstable sampling values caused by UV lamp attenuation, delivers lower detection limit and higher accuracy, and meets sulfur detection requirements for China VI and higher-grade fuels.
Equipped with genuine automatic liquid autosampler with 156 sample positions, enabling unattended analysis with accurate positioning and stable performance.
Mass flow meters are adopted to precisely control gas flow rate, eliminating interference from gas flow and pressure fluctuation to achieve optimal decomposition efficiency and stable, repeatable test results.
Automatic sample/standard switching, automatic cleaning, automatic bubble removal and automatic calibration curve plotting.
Single-point calibration with standard sample rechecking; the high voltage can be adjusted arbitrarily without re-making the calibration curve.
Reaction chamber: Good light-shielding performance with dual sealing design against light leakage and gas leakage (proprietary technology). The light-tight and airtight structure significantly improves detection sensitivity and repeatability, delivering better test results than similar domestic instruments.
Compressed air (from air generator) can be used as reaction gas and carrier gas to replace traditional high-purity oxygen (reaction gas) and high-purity argon (carrier gas), greatly reducing operating cost.
Automatic shutdown protection against unstable gas source pressure to prevent system contamination.
Adopts internationally popular circuit design and imported components to reduce instrument noise and improve detection sensitivity.
Imported high-quality heating furnace filaments for longer service life of the cracking furnace.
Heavy-load high-precision ball-bearing slide rails for fast and smooth movement of the cracking furnace, improving user experience and maintenance convenience.
Uses imported xenon lamp as UV light source instead of traditional zinc lamp, with no extra cost. Its service life exceeds 5 years. Excellent stability guarantees test accuracy and greatly cuts maintenance cost.
Core parts: membrane dryer, optical filter, xenon lamp, photomultiplier tube, sampling needle and other key consumables are original imported from USA, Japan and Australia to guarantee instrument quality and test accuracy.
Equipped with 50 μL imported sampling needle; injection volume is adjustable as required.
Instrument calibration: Calibrated with certified reference